Laboratory water quality analyzer
Beijing JXHS Environmental Protection Science Co., Ltd.

JX-820 multi-parameter water quality analyzer (COD / ammonia nitrogen / total phosphorus / total nitrogen)

JX-820 multi-parameter water quality analyzer (COD / ammonia nitrogen / total phosphorus / total nitrogen) 2019-5-6 This article has been read445 times

JX-820型多参数水质分析仪.jpg

JX-820 multi-parameter water quality analyzer (COD / ammonia nitrogen /  total nitrogen)

Application field

1. COD, ammonia nitrogen (measured by a separate machine).

2. Can be used for the analysis of COD, ammonia nitrogen and total phosphorus.

3. Support DIS-16 type (or optional DIS-25/1A/36) digestion instrument.

4. With printing and USB data upload.

5. For wastewater testing.

Main feature

1.Using imported high-brightness, long-life cold light source and high-stability narrow-band interference optical system, the instrument has high stability and repeatability.

2. The temperature of the digestion instrument is automatically controlled by temperature and timing, and the precision is high.

3. The digestion temperature and time can be set steplessly for other purposes.

4. Digestion and colorimetry use the same digestion colorimetric tube, which has high strength and hardness and can be reused many times.

5. With print function: the test record can be printed immediately or the record print can be queried.

6. With USB interface data upload function, the data record stored by the instrument can be uploaded to the computer for saving or editing and printing.

7.LCD large-screen LCD display, long life and low failure rate, operating Chinese menu display.

The main technical parameters

1. Measurement range: (exceeding the dilution analysis)

COD: 5~2000mg/L

2. Indication error:

COD: ≤±5%

Ammonia nitrogen: ≤ ± 3% and the upper limit of 0.2

3. Repeatability: ≤3%

4. Optical stability: the instrument absorbance value drifts less than 0.002A within 20min

6. Weight: 4Kg

7. Dimensions: 340mm × 250mm × 130mm

8. Power consumption <10W

Digestion instrument

1. Temperature control range: room temperature ~ 180 ° C can be set

COD digestion temperature 165 ° C

2. Temperature control accuracy: ±0.5 °C

3. Digestion time: adjustable from 1-999min.

COD: 15 min

4. Batch volume: 16 water samples (optional 25, 36, 20)

5. Power consumption: <500W

6. Weight: 6kg

7. Dimensions: 216mm × 320mm × 146mm

JX-820 multi-parameter water quality analyzer (COD / ammonia nitrogen / total phosphorus / total nitrogen)
Previous:HS-6010KS type open flash point testernext:JX-9100 reverse osmosis equipment