Laboratory water quality analyzer
Beijing JXHS Environmental Protection Science Co., Ltd.

JX-808 type iron ion analyzer

JX-808 type iron ion analyzer 2019-5-7 This article has been read617 times

JX-808型铁离子分析仪.jpg

JX-808 type iron ion analyzer

Application field

It is widely used in the analysis of ion content in electric power, chemical industry, metallurgy, water treatment, pharmaceutical, industrial wastewater and biochemical water.

Main feature

1, Chinese LCD display, full film touch button.

2. The receiving photocell adopts imported components. The stability and accuracy of the measurement are fully guaranteed.

3. Waterproof case, calibrate with blank water without any reagent. The calibration process is greatly simplified and the work efficiency is improved.

4, unique selection of multi-point curve calibration function. Greater assurance of the accuracy of the instrument over the range.

The main technical parameters

1. Colorimetric method: o-phenanthroline

2, measurement range: 0 ~ 200μg / L

3. Resolution: 0.1μg/L

4. Accuracy: ±1.5% F.S

5, repeatability: no more than 1%

6. Stability: ±1.5% F.S/4h

7, power supply: AC220 ± 22V, 50 ± 1Hz

8, power: 30W

9. Dimensions: 400×180×260 (H×W×D)

10, weight: 4kg

JX-808 type iron ion analyzer
Previous:JX-809 hydrazine analyzernext:JX-9100 reverse osmosis equipment